Although experimental evidence shows that the response of biological systems to UV radiation could deviate from the Bunsen–Roscoe law, the precise operational threshold where this rule fails during air and surface UV disinfection has not been established for standard bacterial models. To define these practical boundaries, the susceptibility of Escherichia coli K12 exposed to 254 nm radiation was evaluated across a representative fluence-rate range (0.0025 to 2.056 Wm−2). Inactivation kinetics for E. coli K12 remained unchanged from 0.0212 to 2.056 Wm−2, demonstrating that the Bunsen–Roscoe law holds over most of the fluence-rate range occurring during air and surface disinfection. However, at 0.0106 Wm−2, a fluence rate comparable to the lower values estimated during disinfection procedures, there was a reduction in lethality, and prolonged exposure resulted in higher bacterial survival at identical total fluences. Reciprocity law validity appears to be limited at the lower end of the range of fluence rates which typically occurs during disinfection procedures. Consequently, to assess the lethal effect of a treatment from exposure time and radiation intensity, it should be verified that the susceptibility of the target microorganism is the same at the maximum and minimum fluence rates that occur within the irradiation device.